Setting the Wakeup Configuration
From the Protocol Set tab (see Figure 8-5) the various wakeup settings may be configured. The
wakeup settings determine the system’s initial configuration at power on or when a new case is
started after loading the pump loop.
Figure 8-5 Wakeup Tab of the Configuration Mode
ATS Mode of Operation
Using the two-part toggle button it is possible to set the wakeup operating mode for intraoperative
and postoperative protocols to either Automatic or 1 Touch. The mode selected with this setting will be
the mode that is active by default every time an intraoperative or postoperative protocol is selected
either when a new case is started or after processing with a sequestration protocol.
If 1 Touch mode is disabled in the Features tab, then it will not be possible to set it as the wakeup
mode here.
This setting does not affect the default mode for sequestration protocols (which is always Automatic).
The Autostart and Continue wakeup settings determine the state of the Autostart and Continue
automations, respectively, whenever an intraoperative or postoperative protocol is activated after a
new case or after processing with a sequestration protocol.
Autostart Levels
Each bowl size (55, 125, 175, and 225 ml) may have a unique wakeup Autostart target volume
configured which will be used as the default Autostart target volume when starting a new case with
the corresponding bowl size. The volume for each bowl size may be set from 200 ml up to 3,500 ml,
with a step of 100 ml.
Vacuum Set Points
Each vacuum mode (Intra and Post-op) may have a unique wakeup negative pressure set (mmHg).
These set points are used every time the vacuum operating mode is changed (including at module
power on).
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...