Note:
These Quick Operating Instructions represent only an
overview of the operating procedures. For a complete setup and
use of the machine, please refer to the Xtra Operator’s Manual.
C-4
20077/021 US
X
TRA
Operator’s Manual
1.
Install the Blood Collection Reservoir:
k. The branch ending with a male port is used to connect the reservoir to the
XTRA Bowl Set, the second one to the Oxygenator Extension Line
2.
Connect the Suction Line:
a. Raise the IV pole and the display completely
2.
Connect the Suction Line:
b. Prepare a bag or bottle of anticoagulant solution by mixing 30,000
international units of heparin per liter of sterile (injectable) normal saline
2.
Connect the Suction Line:
c.
Remove the dual lumen suction line from its package and carefully open the
outer wrap
2.
Connect the Suction Line:
d. Pass the inner sterile wrapped assembly to the sterile field. At the sterile
field, unwrap the assembly, remove the protective cover, and attach the
connector to a suction cannula
2.
Connect the Suction Line:
e. Pass the other end (the split end) of the aspiration line back from the sterile
field to the X
TRA
operator
2.
Connect the Suction Line:
f.
Close the roller clamp on the small bore tubing
2.
Connect the Suction Line:
g. Remove the protective cover on the large bore tubing of the suction line
2.
Connect the Suction Line:
h. Attach the tubing to one of the inlet ports on the lid of the reservoir. The inlet
ports are orange-capped
2.
Connect the Suction Line:
i.
Hang the anticoagulant solution container on one of the holders of the IV
pole
A
B
Quick Start Instructions:
Setting Up the Disposables
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...