Troubleshooting
X
TRA
Operator’s Manual
20077/021 US
14-3
Stop Button
The Stop button is always available regardless of the alarm or warning.
Pressing the Stop button during an alarm or warning will bring the system to the safe Ready Screen,
dismissing the active alarm/warning.
For a further description of the Stop button, refer to “Stop Button” on page 5-55.
Alarms and Warnings During Setup
Table 14-1 lists the warnings which may occur during the setup phase of the X
TRA
operation. Each
warning is listed in alphabetical order by its message text.
If the warning persists after performing the actions listed in the “Operator Responses” column, call the
authorized technical support service.
CAUTION
It is possible to temporarily deactivate some of the controls that the machine
does. The operator can make use of this possibility, under his own
responsibility, therefore directly managing those controls.
Message Text
Warning
or Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management Text
Backup memory failure
Warning
The backup memory is not
functioning.
The machine will continue
working anyway, but it is
highly recommended to call
the authorized technical
service to restore the
backup memory and avoid
future loss of data.
Press X to clear the warning
message and continue.
Backup memory failure
Bowl size not
recognized
Warning
The barcode label cannot
be automatically read by
the sensor.
Manually select the current
bowl size, then press PLAY
to resume.
If the problem persists, call
the authorized technical
service.
Bowl size not recognized
Centrifuge lid open
Warning
The centrifuge lid might be
open.
Close the centrifuge lid and
press PLAY to resume.
If the problem persists,
contact the authorized
technical service.
Centrifuge lid open
Table 14-1 List of Warnings Which Might Occur During Setup
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...