Maintenance
15-2
20077/021 US
X
TRA
Operator’s Manual
Routine Visual Inspection
The X
TRA
equipment should be inspected periodically for any problems such as bent or broken
switches, frayed or twisted power cords, and loose or missing hardware. Discontinue use of the X
TRA
if
the device displays one or more of the above conditions until the problem is corrected and it has been
verified that the device is operating correctly. Verify that the connectors, located on the exterior of the
X
TRA
, for Data and Ethernet are intact and are not damaged and that the cables are correctly
connected.
Check the following components:
• Power supply cable (see Figure 15-1 “A”): check the power supply insulation and shielding is
undamaged along the entire length of the cable (cracks, cuts, clearly visible kinks).
• Vacuum Pump and Serial Connection cables (see Figure 15-1 “B”): check the power supply
insulation and shielding is undamaged along the entire length of the cable (cracks, cuts, clearly
visible kinks).
Figure 15-1 Power Supply Cable (A); Serial Cable and Vacuum Pump Power Supply (B)
If any of the above cables are found damaged replace it. Have the service technician check defective
accessories in any case.
Cleaning and Disinfecting
Apart from the hygienic aspect, it is essential for the operational safety and reliability of the X
TRA
that
it be kept clean. Perform the following cleaning routine every time it is necessary.
Perform the disinfection only in case of loss of blood.
Before cleaning the X
TRA
, disconnect it from the mains power supply and ensure that the system is
switched off.
External Surfaces
Clean all surfaces of the X
TRA
equipment, including the trolley surfaces.
Cleaning
To clean the external surfaces use soapy water, ethyl alcohol and ammonia based cleaning solutions.
Disinfecting
Use disinfectant products specifically provided for rubber / plastic medical tools and devices.
WARNING
Comply with the dilution required and with instructions provided by the
manufacturer of the product (read carefully instructions for use and label of the
product).
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...