
Troubleshooting
14-24
20077/021 US
X
TRA
Operator’s Manual
Electronic Failure - Memory or
Microcontroller
E3
E16
E26
E27
E28
E30
E36
E40
E61
E87
• Switch the machine OFF.
• Wait a few seconds, switch the machine on and
touch the “Retain” button.
Electronic Failure - Serial Communication
E4
E5
E6
E8
E9
E10
E12
E13
E14
E31
E32
E39
• Switch the machine OFF.
• Wait a few seconds, switch the machine on and
touch the “Retain” button.
Electronic Failure - NMC Board
Overtemperature
E23
• Switch the machine OFF.
• Wait a few minutes to allow the board to cool
down, then switch the machine on and touch the
“Retain” button.
Pump Loop Ejector Failure
E123
• Switch the machine OFF.
• Wait a few minutes, switch the machine on and
touch the “Retain” button.
Error Family
Error Service
Code(s)
Operator Response
Table 14-3 List of Fatal Error Families (Continued)
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...