Note:
These Quick Operating Instructions represent only an
overview of the operating procedures. For a complete setup and
use of the machine, please refer to the Xtra Operator’s Manual.
X
TRA
Operator’s Manual
20077/021 US
C-9
1.
Install the disposable set. Draw 200 ml of anticoagulant
solution into the blood collection reservoir prior to aspirating
blood.
Prime the blood administration set, if
applicable.
2.
Turn on the machine using the ON/OFF switch located on the
rear panel. Once the system has booted and the Setup
Screen is displayed, touch the Load button (A) to autoload
the pump loop tubing and begin a new case.
3.
Looking at the status area (A) of the Ready Screen, make
sure that the active protocol is one of the ATS protocols (Pstd,
Popt, or Post-op), the active operating mode is either 1 Touch
or Automatic, and that the Autostart and Continue icons are
present.
If they aren’t, touch the Menu button (B) to enter the Menu
Screen, select the Protocols/Mode tab from the left of the
Menu Screen, set the active protocol to an ATS protocol, set
the mode to 1 Touch or Automatic, and ensure the Autostart
and Continue automations are enabled.
4.
Once ready, touch the Start button (A) to begin processing.
Alternatively, wait until there is sufficient blood in the
reservoir to trigger the Autostart automation (1,600 ml by
default).
5.
The machine will begin processing through the
Fill-Wash-Empty cycle, starting a new cycle at the end of each
Empty phase, until the reservoir is empty.
6.
When the reservoir is empty, the “Reservoir empty. Bowl not
filled” alarm is issued. Touch the Last Bowl button (A) to
concentrate and wash the remaining blood and empty the
RBC line.
7.
The case is complete. Proceed in one of the following ways:
• Save the case tally data to an external destination using the “Select destination” drop-down menu and the Send Output button (A).
• To process another case, touch the Start New Case button (B) to return to the Ready Screen and begin a new case.
• If there are no other cases to process, shut down the machine.
A
B
A
A
A
A
B
Quick Start Instructions:
Running an A
T
S Case
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...