Preoperative Sequestration (PPP and PRP)
10-20
20077/021 US
X
TRA
Operator’s Manual
9.
Connect the suction line, described in “Step 2: Connecting the Suction Line” on page 4-4.
10. See
for instructions on processing for intraoperative blood recovery.
Tally Information
A table containing total processed and collected volumes is available at any time during processing
from the Tally tab of the Menu Screen (see Figure 10-12).
Figure 10-12 Tally Tab of the Menu Screen (PRP2)
For a detailed explanation of the
Tally tab of the Menu Screen,
read “Tally Tab” on page 5-44.
The Tally tab of the Menu Screen displays the total volume of whole blood processed in the second
column, followed by the total volumes of RBC, PPP, and PRP collected in each cycle. The totals at the
bottom of each column reflect the volumes of the entire case.
The PPP and PRP volumes are completely separate from the intraoperative blood recovery volumes.
The tallies are reset only when the
X
TRA
is shut down or a new case is started.
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...