
System Description
X
TRA
Operator’s Manual
20077/021 US
3-9
Centrifuge Assembly
The centrifuge assembly is located at the front of the machine and includes the centrifuge arm,
centrifuge plate, and centrifuge lid. The centrifuge plate rotates at 5,600 rpm during processing and is
not operator adjustable during the autotransfusion protocols. This speed produces rapid separation of
the red cells with minimal trauma. The centrifuge speed can be manually adjusted by the operator
only during PPP and PRP protocols (from 5,600 down to 2,400 rpm).
Figure 3-13 Centrifuge Assembly
The centrifuge assembly (see Figure 3-13) consists of:
1.
The (1) centrifuge arm and the position sensor, which senses the correct positioning and locking
of the bowl head.
2.
The (2) centrifuge housing, which contains possible leakages of fluids from the bowl.
3.
The (3) centrifuge housing light.
4.
The fluid loss sensor (not visible in figure), which detects fluid leakages in the centrifuge and
stops the centrifuge and pump in case of a leak.
5.
The two RBC (Buffy-Coat) sensors (not visible in figure), high and multipoint, which sense when
the red blood cells have reached a predetermined level within the centrifuge bowl.
6.
The centrifuge lid (not visible in figure) that automatically latches and remains locked as a safety
precaution until the centrifuge and pump come to a complete stop.
Note:
The X
TRA
machine prevents access to the centrifuge (while in operation), through a latched
lid. In the event of a power failure, this latch will open.
CAUTION
If the values regarding volumes, hematocrit, supernatant removal and waste
color line are important for the patient, it will always be necessary to use
other standard measurement instruments of the hospital.
WARNING
2
1
3
Summary of Contents for 75220
Page 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Page 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Page 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Page 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Page 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Page 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Page 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Page 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Page 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Page 282: ...A 6 20077 021 US XTRA Operator s Manual...
Page 284: ...B 2 20077 021 US XTRA Operator s Manual...
Page 286: ...C 2 20077 021 US XTRA Operator s Manual...
Page 298: ...D 2 20077 021 US XTRA Operator s Manual...