Remote Control Commands
R&S
®
SGU100A
193
User Manual 1176.7726.02 ─ 09
be completed automatically after starting the calibration process. Otherwise you have
to follow the steps described in
"Calibration process of the R&S SGU"
the given example, replace the
:CALibration:xxx:STARt
SCPI command
with
:TEST:ALL:STARt
and
:CALibration:xxx:RESult
with
:TEST:ALL:RESult
. We assume that a remote PC is connected to the instru-
ments, the remote PC and the instruments are switched on and a connection between
them is established.
The self tests returns a "0" if the test is performed successfully, otherwise a value other
than "0" is returned. None of the commands of this system has a
*RST
value.
Improper use can destroy the assembly
The respective hardware assembly responds directly to the
:TEST:DIRect
command;
any safety mechanisms are bypassed. The command is intended for servicing purpo-
ses and should be used only by the Rohde
&
Schwarz service personnel.
............................................................................................... 193
:TEST:ALL:STARt
Starts a selftest. Use the command
Use
to query for settings required by the instrument.
Usage:
Event
:TEST:ALL:RESult?
Queries the result of the performed selftest. Start the selftest with
.
Return values:
<Result>
0 | 1 | RUNning | STOPped
*RST:
STOPped
Example:
See
Chapter 11.1.1, "Performing General Tasks for Instrument
Usage:
Query only
:TEST:KEYBoard[:STATe]
<State>
Enables/disables the keyboard and the LED test state.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
TEST Subsystem