Remote Control Commands
R&S
®
SGU100A
155
User Manual 1176.7726.02 ─ 09
See
"Calibration process of the R&S SGU"
on page 151 for a step by step description
of the calibration process.
Usage:
Event
:CALibration:LEVel:RESult?
Queries the result of the internal level adjustment. Start the adjustment with the SCPI
command
.
Return values:
<Result>
0 | 1 | RUNning | STOPped
Usage:
Query only
:CALibration:LEVel:TEMPerature?
Queries the delta temperature since the last level adjustment.
Return values:
<Temperature>
string
Usage:
Query only
11.6
CONNector Subsystem
.........................................................................................155
:CONNector:TRIGger:OMODe
<Mode>
Sets the operating mode of the trigger connector.
Parameters:
<Mode>
SVALid | SNValid | PEMSource
SVALid|SNValid
signal valid /not valid
PEMSource
external pulse modulator source
Manual operation:
See
11.7
DIAGnostic Subsystem
...........................................................................................156
.......................................................................................156
DIAGnostic Subsystem