General Instrument Settings and Instrument Setup
R&S
®
SGU100A
101
User Manual 1176.7726.02 ─ 09
:SYSTem:COMMunicate:USB:RESource?
:SYSTem:COMMunicate:PCIexpress:RESource?
Device Identity
Triggers the device identification function. The [LAN] LED on the front panel of the
selected instrument blinks.
See also
Chapter 6.4.4, "Bidirectional Instrument Identification "
8.9
Factory Preset
The "Factory Preset" dialog provides a function to reset the instrument's settings to
their factory states
1. To access this dialog, select "SGMA-GUI > Instrument Name > Setup > Factory
Preset".
2. Select "Execute Factory Preset".
The instrument's settings are reset to their factory states. Security settings and
user data are not effected.
Execute Factory Preset
Reset the instrument's settings to their factory state.
Note:
Because "Factory Preset" resets the "Remote Channel Settings" and "Network
Settings" to the default values, executing factory preset via remote control may termi-
nate the connection to the instrument, if these settings had been configured to values
different to the default ones!
The factory preset function resets nearly all instrument settings. In addition to the regu-
lar preset, a "Factory Preset" resets also the following values:
●
Power on settings ("Level" dialog)
●
Network settings including hostname ("Setup > Network Setting" dialog)
●
Remote Channel settings ("Setup > Remote Channel" dialog)
●
Eco Mode state ("Setup > Eco Mode" dialog)
To maintain security, password settings and all settings protected by these passwords
like disabled USB and LAN connections are not changed.
Not affected by the "Factory Preset" are also user data, lists or instrument settings
files, created, for example, with the "File Save As" function.
Remote command:
Factory Preset