Remote Control Commands
R&S
®
SGU100A
186
User Manual 1176.7726.02 ─ 09
:SYSTem:SERRor?
Returns a list of all errors existing at the time when the query is started. This list corre-
sponds to the display on the info page under manual control.
Return values:
<Serror>
string
Example:
SYST:SERR
queries all errors existing in the error queue.
Example:
Response: -221, 'Settings conflict', 153,
'Input voltage out of range'
The two returned errors have occurred since the error queue
was last queried.
Usage:
Query only
:SYSTem:VERSion?
Queries the SCPI version the instrument's command set complies with.
Return values:
<Version>
string
Example:
SYST:VERS
queries the SCPI version.
Response:
"1996"
The instrument complies with the SCPI version from 1996.
Usage:
Query only
:SYSTem:COMMunicate:NETWork:IPADdress
<IpAddress>
Sets the IP address.
Parameters:
<IpAddress>
string
Range:
0.0.0.0. to ff.ff.ff.ff
Example:
SYSTem:COMMunicate:NETWork:IPADdress "7.8.9.10"
sets the IP address of the instrument.
Manual operation:
See
:SYSTem:COMMunicate:NETWork:IPADdress:MODE
<Mode>
Selects manual or automatic setting of the IP address.
Parameters:
<Mode>
AUTO | STATic
*RST:
n.a. (factory preset: AUTO)
SYSTem Subsystem