Remote Control Commands
R&S
®
SGU100A
154
User Manual 1176.7726.02 ─ 09
See
"Calibration process of the R&S SGU"
on page 151 for a step by step description
of the calibration process.
Usage:
Event
:CALibration:IQModulator:RESult?
Queries the result of the internal I/Q adjustment. Start the adjustment with the SCPI
command
:CALibration:IQModulator:STARt
.
Return values:
<Result>
0 | 1 | RUNning | STOPped
Usage:
Query only
:CALibration:IQModulator:TEMPerature?
Queries the delta temperature since the last adjustment of the IQ modulator.
Return values:
<Temperature>
string
Usage:
Query only
:CALibration:IQModulator:LOCal:STARt
Starts an adjustment for the I/Q modulator of the current frequency. The I/Q modulator
is adjusted with respect to carrier leakage, I/Q imbalance and quadrature. Query the
result of the adjustment with the SCPI command
.
See
"Calibration process of the R&S SGU"
on page 151 for a step by step description
of the calibration process.
Usage:
Event
:CALibration:IQModulator:LOCal:RESult?
Queries the result of the I/Q modulation adjustment for the current frequency. Start the
adjustment with the SCPI command
:CALibration:IQModulator:LOCal:STARt
Return values:
<Result>
0 | 1 | RUNning | STOPped
Usage:
Query only
:CALibration:LEVel:STARt
Starts an internal level adjustment. Query the result of the adjustment with the SCPI
command
.
CALibration Subsystem