Instrument Tour
R&S
®
SGU100A
30
User Manual 1176.7726.02 ─ 09
Pressing the [ID] key while the instrument is active, opens the "SGMA-GUI > Setup >
Instruments > Configure Instruments > Edit Instrument" dialog of the corresponding
instrument on the remote controller.
Pressing the ID key of an inactive instrument and starting "SGMA-GUI > Setup >
Instruments > Scan", leads to an automatic activation of the instrument in the SGMA-
GUI.
3.2
Rear Panel Tour
This section provides an overview of the connectors on the rear panel of the instru-
ment. For technical data of the connectors, refer to the data sheet.
Figure 3-2: Rear panel view
Risk of damaging components
Always use an appropriate 8 mm torque wrench to tighten SMA connectors. Limit the
tightening torque to 60 Ncm. Never use an open-end wrench.
[PCI Express Connector]
The PCIe (Peripheral Component Interconnect Express) single lane interface allows
remote control with optimized speed.
For details, see
Chapter 2.3.3, "Connecting a Controller via PCI Express"
[USB Connector]
The USB (universal serial bus) interface, type micro, allows you to connect various
external devices, e.g.:
●
If the interface is configured as a device interface, the R&S
SGU can be connected
to other devices like a remote computer. On the computer, you can perform a firm-
Rear Panel Tour