Remote Control Commands
R&S
®
SGU100A
149
User Manual 1176.7726.02 ─ 09
EXTension
The instrument is controlled by another Rohde&Schwarz instru-
ment. This parameter is used by the SGMA-GUI to disable set-
tings when the instrument is in extension mode.
*RST:
STDalone
:LOCK?
<Lock Request Id>
Sends a lock request ID which uniquely identifies the controller of the instrument.
Parameters:
<Lock Request Id>
Number
0
test query to check whether the instrument is locked
Controller ID
request lock from the controller with the specified Controller ID
Return values:
<Value>
Number
0
request refused; the instrument is already locked to another
<Lock Request Id>, i.e. to another controller
1
request granted
Example:
:LOCK? 12345
Response: 1
:UNL 12345
Usage:
Query only
Manual operation:
See
:UNLock
<Unlock Id>
Unlocks an instrument locked to a controller with Controller ID =
<Unlock Id>
.
Setting parameters:
<Unlock Id>
Number
Unlock ID which uniquely identifies the controller of the instru-
ment. The value must match the Controller ID
<Lock Request
Id>
.
Usage:
Setting only
Manual operation:
See
General Commands