First Steps with the Instrument
R&S
®
SGU100A
34
User Manual 1176.7726.02 ─ 09
d) Connect the [RF OUT] of the R&S SGU to the [RF IN] of the signal analyzer.
2. Switch on the R&S SGS and the R&S SGU.
The [POWER ON/STANDBY] keys have to be
green
and not blinking.
3. Press the [ID] keys on the front panels of the R&S SGS and the R&S SGU (only
required if the components are connected via a company network).
4. On the connected remote PC, start the R&S SGMA-GUI software application.
The main panel of the application and the configure instruments dialog open. Both
instruments are added automatically to the instruments list and to the main panel of
the R&S SGMA-GUI software.
The main panel provides a quick access to the main settings of the configured and
activated instruments. The display shows one row per instrument with the instru-
ment specific settings. The rows comprise the instrument, the connection state, the
used frequency and power level and the state of the RF output and the modulator.
5. In the R&S SGMA-GUI main panel, the green indicator in front of the instrument's
name confirms that there is a connection between the instrument and the remote
PC and that the instrument is recognized by the software.
6. In the R&S SGMA-GUI main panel, select the rows corresponding to the corre-
sponding instruments to be configured and select "Instrument Name > Preset" to
restore their predefined settings.
7. Select "SGMA-GUI main panel > R&S SGS > Extension".
The "Extension" dialog opens.
R&S SGU as an Extension to the R&S SGS