System Overview
R&S
®
SGU100A
44
User Manual 1176.7726.02 ─ 09
Figure 5-5: Connection of an R&S SGS and an R&S SGU in a company network
1. Connect the test equipment as shown in
a) Connect the R&S SGS, the R&S SGU and the controller to the company net-
work.
b) Connect the [RF OUT] of the R&S SGS to the [LO IN] of the R&S SGU.
2. Switch on the R&S SGS and the R&S SGU.
The [POWER ON/STANDBY] keys have to be
green
and not blinking.
3. Press the [ID] keys on the front panels of the R&S SGS and the R&S SGU.
4. On the connected remote PC, start the R&S SGMA-GUI software application.
The main panel of the application and the configure instruments dialog open. Both
instruments are added automatically to the instruments list and to the main panel of
the R&S SGMA-GUI software.
Note:
If you connect the instruments to the company network for the first time this
process may take several minutes.