Remote Control Commands
R&S
®
SGU100A
176
User Manual 1176.7726.02 ─ 09
[:SOURce]:POWer:SCHaracteristic
<Characteristic>
Selects the characteristic for the level setting.
Parameters:
<Characteristic>
AUTO | UNINterrupted | CVSWr | USER | MONotone
UNINterrupted
: Uninterrupted Level setting,
CVSWr
: Con-
stant-VSWR,
MONotone
: Strictly Monotone
*RST:
AUTO
Example:
Chapter 11.1.3, "Advanced Task for Optimizing Perform-
Manual operation:
See
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude]
<Amplitude>
Sets the RF level at the RF output connector of the instrument.
Parameters:
<Amplitude>
float
Range:
-120 to 25
Increment: 0.01
*RST:
-10
Example:
See
Chapter 11.1.3, "Advanced Task for Optimizing Perform-
Manual operation:
See
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet
<Offset>
Specifies the constant level offset of a downstream attenuator/amplifier. If a level offset
is entered, the level entered with
no longer corresponds
to the RF output level.
The following correlation applies:
:POWer
= RF output level +
POWer:OFFSet
.
Entering a level offset does not change the RF output level, but rather the query value
of
:POWer
.
Only dB is permitted as the unit here. The linear units (V, W, etc.) are not permitted.
Parameters:
<Offset>
float
Range:
-100 to 100
Increment: 0.1
*RST:
0
Manual operation:
See
SOURce:POWer Subsystem