Performing Configuration Tasks
R&S
®
SGU100A
111
User Manual 1176.7726.02 ─ 09
2. On the PC, open a windows explorer window.
3. To connect to the R&S
SGU, enter the name of the instrument or its IP address in
the windows taskbar.
4. Enter the user name and password to connect to the R&S
SGU. The default user
name is
instrument
and the password is
instrument
.
A folder opens, containing the
share
and the
update
folder.
5. Open the
update
folder and copy the new firmware update file in it.
The update starts automatically.
Unsuccessful or erroneous firmware update
An erroneous or unsuccessful installation of firmware update package is indicated by a
combination of one orange and red LEDs on the front panel.
Refer to the service manual for a description of the displayed error code or contact the
customer support center.
9.6
How to Activate Options
A firmware update before the activation of the SW option may be required.
Refer to the description of the SW option for the required firmware version.
See also
Chapter 9.5, "How to Install a New Firmware Version on the Instrument"
on page 110 for instruction on how to update the firmware version.
1. Select "SGMA-GUI main panel > Instrument Name > Setup > Install SW-Options".
2. Select "Option Key" and enter the key code delivered with the new option.
The new option is now enabled and ready for operation.
9.7
How to Manually Set a PCIe Direct Connection
between an R&S SGS and an R&S SGU
To build a direct PCIe connection between an R&S SGU and an R&S SGS, which has
a "Controller > Revision" < 5, first you have to set the correct PCIe interface mode
manually. If your R&S SGS has a "Controller > Revision" 5 or higher, these settings are
done automatically.
You can check the "Controller > Revision" of your instrument in the "SGMA-GUI >
Instrument Name > Hardware Config" dialog.
How to Manually Set a PCIe Direct Connection between an R&S SGS and an R&S SGU