Remote Control Commands
R&S
®
SGU100A
168
User Manual 1176.7726.02 ─ 09
Example:
See
Chapter 11.1.2, "Generating an I/Q Modulated Signal"
Manual operation:
See
:OUTPut[:STATe]:PON
<Mode>
Selects the state which the RF output takes when the instrument is switched on.
Parameters:
<Mode>
OFF | UNCHanged
OFF
When the instrument is switched on, the output is deactivated .
UNCHanged
When the instrument is switched on, the output remains in the
same state as it was before the instrument was switched off.
*RST:
UNCHanged
Example:
See
Chapter 11.1.2, "Generating an I/Q Modulated Signal"
Manual operation:
See
:OUTPut:AFIXed:RANGe:LOWer?
Queries the minimum level which can be set without the attenuator being adjusted.
Return values:
<Lower>
float
Default unit: dBm
Example:
See
Chapter 11.1.3, "Advanced Task for Optimizing Perform-
Usage:
Query only
Manual operation:
See
:OUTPut:AFIXed:RANGe:UPPer?
Queries the maximum level which can be set without the attenuator being adjusted.
Return values:
<Upper>
float
Default unit: dBm
Example:
See
Chapter 11.1.3, "Advanced Task for Optimizing Perform-
Usage:
Query only
Manual operation:
See
OUTPut Subsystem