Remote Control Commands
R&S
®
SGU100A
153
User Manual 1176.7726.02 ─ 09
:CALibration:IQModulator:LOCal:STARt
........................................................................... 154
:CALibration:IQModulator:LOCal:RESult?
.........................................................................154
...............................................................................................154
............................................................................................ 155
:CALibration:LEVel:TEMPerature?
................................................................................... 155
:CALibration:HREQuest?
Queries the settings required from the instrument during a calibration or a selftest.
See
"Calibration process of the R&S SGU"
on page 151 for a step by step description
of the calibration process.
Return values:
<Command>
string
All of the commands described below return the values in the
parameter's default unit.
:VERsion <Number>
Shows the command set version with which the instrument is
compliant
:OUTP:STATe <Number>
Shows the state required of the local oscillator
:SOUR:FREQ <Number>
Shows the frequency required of the local oscillator
:SOUR:POW<Number>
Shows the power required of the external local oscillator
:PROG:CREate <Text>
Asks the controller to create a progress bar
:PROG:REName <Text>
Asks the controller to rename the progress bar
:PROG:REMove
Asks the controller to delete the most recently created progress
bar
:PROG:UPDate <Number>
Shows the current progress value in percent
:NOP/:NO DATA
Intermediate message. No action is required
:CAL:HREQ:STOP
End of the calibration procedure
Example:
See
Usage:
Query only
:CALibration:IQModulator:STARt
Starts an internal I/Q adjustment. Query the result of the adjustment with the SCPI
command
:CALibration:IQModulator:RESult?
.
CALibration Subsystem