![Rohde & Schwarz 1424.7805K02 User Manual Download Page 67](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791067.webp)
Measurement and display configuration
R&S
®
NRX
67
User Manual 1178.5566.02 ─ 09
Remote command:
TRIGger<undef>:ALL:SLOPe
TRIGger<Measurement>[:CHANnel<Channel>]:SLOPe
Level ← Trigger Advanced
See
Delay ← Trigger Advanced
Sets the delay between the trigger event and the beginning of the actual measurement.
Remote command:
TRIGger<undef>:ALL:DELay[:VALue]
TRIGger<Measurement>[:CHANnel<Channel>]:DELay[:VALue]
Dropout ← Trigger Advanced
If the trigger level has been underrun with internal triggering, it is sometimes useful to
wait some time (drop-out time) before retriggering. The measurement of unwanted sig-
nal components can thus be suppressed. See
Chapter 6.1.3.3, "Dropout time"
Remote command:
TRIGger<undef>:ALL:DTIMe
TRIGger<Measurement>[:CHANnel<Channel>]:DTIMe
Holdoff ← Trigger Advanced
Sets the hold-off time, see
Chapter 6.1.3.4, "Hold-off time"
Remote command:
TRIGger<undef>:ALL:HOLDoff
TRIGger<Measurement>[:CHANnel<Channel>]:HOLDoff
Hysteresis ← Trigger Advanced
Sets the hysteresis. A trigger event occurs, if the trigger level:
●
Falls below the set value on a rising slope.
●
Rises above the set value on a falling slope.
Thus, you can use this setting to eliminate the effects of noise in the signal for the edge
detector of the trigger system.
Remote command:
TRIGger<undef>:ALL:HYSTeresis
TRIGger<Measurement>[:CHANnel<Channel>]:HYSTeresis
Specific Trigger
Not available for each sensor type. Groups the specific trigger settings.
Jitter Suppression ← Specific Trigger
Defines the method how to cope with the misalignment between the trigger event and
the sample point.
"Compensate"
Compensation means resampling of trace result.
"Measure"
Does not perform resampling, but stores the measured trigger jitter.
Configuration for all measurement types