CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage:TCONtrol:AUTO
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage:TCONtrol[:ENUM]
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage[:STATe]
CALCulate<Measurement>[:CHANnel<Channel>][:POWer]:NCORrection[:STATe]
CALCulate<Measurement>[:CHANnel<Channel>][:POWer]:VBWidth:ENUM
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution
<resolution>
Defines the number of significant places for linear units and the number of decimal pla-
ces for logarithmic units which should be free of noise in the measurement result.
The setting is only taken into account, if:
●
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:
TYPE
RES
●
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:
AUTO[:STATe]
ON
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<resolution>
Indicates the number of significant digits.
1
1
2
0.1
3
0.01
4
0.001
Range:
1 to 4
*RST:
3
[SENSe<Sensor>:]AVERage:RESet
Effective for continuous average, burst average, time gate measurements.
Clears the filter buffer.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Event
Manual operation:
See
Configuring sensors