Measurement and display configuration
R&S
®
NRX
83
User Manual 1178.5566.02 ─ 09
Trigger
Gives quick access to selected trigger settings. See
6.5.3
Pulse Analysis dialog
Access: Tap the measurement results that are displayed below the trace.
On the "Time" and "Power" tabs, select the measurement results that are displayed
below the trace. The R&S NRX can display a maximum of 12
measurement results. If
you select more, a warning is displayed.
100
Amplitude
(%)
Time (s)
0
1
0
2
Pulse width
Pulse period
Overshoot
Pulse top
Pulse base
Undershoot
Rise time
Fall time
Figure 6-5: Main pulse analysis parameters and characteristic values
................................................................................................ 85
................................................................................................ 86
Pulse analysis