Remote control commands
R&S
®
NRX
311
User Manual 1178.5566.02 ─ 09
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
[SENSe<Sensor>:]STATistics:OFFSet[:TIME]
<time>
Sets the start of the sampling window, referenced to delay set by
TRIGger<Measurement>[:CHANnel<Channel>]:DELay[:VALue]
.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
0.0 to 10.0
*RST:
0.0
Default unit: s
[SENSe<Sensor>:]STATistics:PEAK?
Queries the peak power value calculated during a CCDF or PDF statistics measure-
ment.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
CALCulate<Measurement>:STATistics:POWer:AVG:DATA?
Effective for statistics measurements.
Queries the average power value in the sampling window.
The command is not fully R&S
NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S
NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Measurement settings and results