![Rohde & Schwarz 1424.7805K02 User Manual Download Page 274](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791274.webp)
Remote control commands
R&S
®
NRX
274
User Manual 1178.5566.02 ─ 09
DISPlay[:WINDow<Window>]:TRACe:MARKer<Marker>:MODE
<mode>
Enables or disables the marker. Also defines the appearance of the marker.
Suffix:
<Window>
.
1 to 4
Measurement pane
<Marker>
1 to 4
Marker (M1 to M4)
Parameters:
<mode>
OFF | RULer | MEASure
OFF
Disables the marker.
RULer
Shows a line at the marker position. Useful if you use the marker
as reference marker,
DISPlay[:WINDow<Window>]:TRACe:
MEASure
Shows a triangle at the marker position.
*RST:
OFF
Manual operation:
See
DISPlay[:WINDow<Window>]:TRACe:MARKer<Marker>:POSition:MODE
<mode>
Defines where the marker is placed.
Suffix:
<Window>
.
1 to 4
Measurement pane
<Marker>
1 to 4
Marker (M1 to M4)
Parameters:
<mode>
FTIMe | FPOWer | RPOSition | RPOWer | RPLeft | RPRight |
PSEarch | MSEarch | RPSLeft | RPSRight | RMSLeft |
RMSRight
FTIMe
At a fixed time, set by the marker position.
FPOWer
At a fixed power value, set by the marker position.
RPOSition
At a time difference of the marker position to the x-position of the
reference marker.
RPOWer
At a power difference of the marker position to the y-position of
the reference marker.
Measurement settings and results