![Rohde & Schwarz 1424.7805K02 User Manual Download Page 108](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791108.webp)
Measurement and display configuration
R&S
®
NRX
108
User Manual 1178.5566.02 ─ 09
Evaluate
Opens the "Evaluate" dialog to configure the sampling window.
"Gate 1" / "Gate 2" / "Gate 3" / "Gate 4"
Select the gate that you want to configure and use for the measure-
ment.
Continuous
Sets unsynchronized acquisition. Set the duration of the sampling
window using
Remote command:
CALCulate<Measurement>:STATistics:TGATe:SELection
t0 / t1 / t2 / t3
Available if
is disabled.
The same gates are used in the time gate and statistics measurements. See
6.9
NRT
Requires the interface for R&S NRT-Z sensors (R&S NRX-B9), see
Used for power reflection measurements with the R&S
NRT
‑
Zxx directional power sen-
sor.
The R&S
NRT
‑
Zxx directional power sensor measures the forward and reverse power.
The forward power is the power flux from the source to the load.
To observe during the power reflection measurement
►
CAUTION!
Risk of electric shock and severe skin burns. During the measurement,
the RF power flow can be high.
After switching on the RF power, do not touch the RF ports.
Never exceed permissible continuous power that is shown in the diagram on the
back of the power sensor.
Further information:
●
Chapter 7.5, "NRT measurement type"
●
Chapter 3.1.8.2, "Optional interface for R&S NRT-Z sensors (R&S NRX-
6.9.1
NRT result display
The R&S NRX displays the forward and reverse power simultaneously.
NRT