Remote control commands
R&S
®
NRX
344
User Manual 1178.5566.02 ─ 09
Enables or disables the automatic termination control. See also
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:TCONtrol[:
ENUM]
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<state>
*RST:
ON
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:TCONtrol[:ENUM]
<mode>
Effective for continuous average, burst average, pulse analysis, timeslot, statistics
measurements.
Defines how the measurement results are output. This is called termination control.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<mode>
MOVing | REPeat
If you use
instead of
MOVing
Outputs intermediate values to facilitate early detection of
changes in the measured quantity. In the settled state, that
means when the number of measurements specified by the
average count has been performed, a moving average is output.
REPeat
Specifies that a measurement result is not output until the entire
measurement has been completed. This means that the number
of measurement cycle repetitions is equal to the set average
count. If the average count is large, the measurement time can
be very long.
The average count is set using
CHANnel<Channel>]:AVERage:COUNt[:VALue]
on page 374.
*RST:
MOVing
Manual operation:
See
Configuring sensors