Remote control commands
R&S
®
NRX
261
User Manual 1178.5566.02 ─ 09
Parameter
Description
<resolution>
Optional. Limit up to which the measurement result probably is free of
noise.
[SENSe<Sensor>:]AVERage:COUNt:AUTO:
<source_list>
Mandatory. Primary and secondary sensor. The number for each sensor
is preceded by the character
@
. The entire expression is enclosed in
parentheses.
Example: (
@
3),(
@
2)
Sensor
C is the primary sensor, and sensor
B is the secondary sensor.
CONFigure<Measurement>[:SCALar][:POWer]:BURSt?
<dtolerance>,
<start_exclude>, <end_exclude>[, <expected_value_or_source_list>,
<resolution_or_source_list>, <source_list>...]
FETCh<Measurement>[:SCALar][:POWer]:BURSt?
<dtolerance>, <start_exclude>,
<end_exclude>[, <expected_value_or_source_list>,
<resolution_or_source_list>, <source_list>...]
READ<Measurement>[:SCALar][:POWer]:BURSt?
<dtolerance>, <start_exclude>,
<end_exclude>[, <expected_value_or_source_list>,
<resolution_or_source_list>, <source_list>...]
MEASure<Measurement>[:SCALar][:POWer]:BURSt?
<dtolerance>,
<start_exclude>, <end_exclude>[, <expected_value_or_source_list>,
<resolution_or_source_list>, <source_list>...]
Power measured by one power sensor.
Suffix:
<Measurement>
.
1 to 8
Measurement
Query parameters:
<dtolerance>
Default unit: s
<start_exclude>
Default unit: s
<end_exclude>
Default unit: s
<expected_value_or_source_list>
<numeric> | <expr>
<resolution_or_source_list>
<numeric> | <expr>
<source_list>
<expr>
For the parameter descriptions, see
ing commands for burst average measurements"
Usage:
Query only
Measurement settings and results