![Rohde & Schwarz 1424.7805K02 User Manual Download Page 96](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791096.webp)
Measurement and display configuration
R&S
®
NRX
96
User Manual 1178.5566.02 ─ 09
resolution is determined by the sampling rate of the power sensor. An external trigger
signal or internal signal triggering is required for synchronization.
Trigger level
Power Trigger event
Timeslot 1
Timeslot 2
Timeslot 3
Timeslot width
Exclude from start
Exclude from end
Start of fence
Length of fence
Time
Figure 6-9: Timeslot measurement parameters
Adopt the timeslot width to the test signal. You can restrict the measurement to the rel-
evant part of a timeslot by excluding intervals at the start and the end of the timeslot.
Also, you can define an exclusion interval, a so-called fence, within the timeslot.
.............................................................................................96
..................................................................................99
6.7.1
Timeslot result display
You can choose between a scalar or graphical result display.
To change the display format
1. Select the timeslot measurement type:
"Measurement Settings"
> "Measurement Type"
> "Timeslot"
2. Select the display format:
"Display"
> "Display Format"
Timeslot