Measurement and display configuration
R&S
®
NRX
55
User Manual 1178.5566.02 ─ 09
"Scalar Analog"
Numeric format with bar chart
"Graphical"
Available for time gate, timeslot measurements.
Measured values are plotted over time.
Remote command:
Auxiliary Values
Available for the graphic displays of continuous average, burst average measure-
ments. Only displayed if the measurement result display is shown in full screen.
Determines which additional information about the measured values is shown in the
display. If you press [1 Trig / Delete], you reset the auxiliary values.
"None"
No additional values are measured.
"Extremes"
Displays the maximum, the minimum and the max-min values since
the search for extreme values has been started. With logarithmic
units, the peak-to-peak value equals the quotient of the measured
values converted into linear units.
"Statistics"
Displays the long-term mean, the standard deviation and the total
number of measurement results that have been evaluated since the
search for statistic values has been started.
Remote command:
Scaling
Configures the display scaling. The available parameters depend on the following set-
tings:
●
●
Scale Lower Limit ← Scaling
If
is set to "Scalar Analog", available for continuous average, burst
average, time gate, timeslot measurements.
Defines the lower limit of the bargraph display.
Remote command:
CALCulate<Measurement>:METer<DirectionalChannel>:LOWer[:DATA]:
RATio:RCOefficient
CALCulate<Measurement>:METer<DirectionalChannel>:LOWer[:DATA]:
RATio:RFRatio
CALCulate<Measurement>:METer<DirectionalChannel>:LOWer[:DATA]:
RATio:RLOSs
CALCulate<Measurement>:METer<DirectionalChannel>:LOWer[:DATA]:
RATio:SWR
CALCulate<Measurement>:METer<DirectionalChannel>:LOWer[:DATA]:
RATio[:VALue]
CALCulate<Measurement>:METer<DirectionalChannel>:LOWer[:DATA][:
POWer]
Configuration for all measurement types