Remote control commands
R&S
®
NRX
329
User Manual 1178.5566.02 ─ 09
Channel calculation func-
tion
<expression>
Description
Reflection coefficient
"RCO(SENSn,SENSm)"
Reflection coefficient of a DUT. No output
unit.
SENSm/SENSn
Sensor
n measures the forward power of a
wave.
Sensor
m measures the reflected power.
Return loss
"RLOS(SENSn,SENSm)"
Return loss of a DUT. The output unit is dB.
SENSm/SENSn
SENSm/SENSn
10 x log
Sensor
n measures the forward power of a
wave.
Sensor
m measures the reflected power.
Reflection ratio
"REFL(SENSn,SENSm)"
Reflection ratio of a DUT. The output unit is
percent.
SENSn
SENSm
100
Sensor
n measures the forward power of a
wave.
Sensor
m measures the reflected power.
With n and m = 1, 2, 3, 4 or A, B, C, D
Table 13-6: Additional <expression> parameters for queries
<expression>
Description
OFF
No channel calculation function is selected.
-
For
SENSn
or
SENSm
, if no primary or secondary sensor is selected.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<expression>
See
. The unit is set by
UNIT<Measurement>:POWer[:VALue]
The query returns two additional expressions, see
.
*RST:
Depends on the selected channel.
Default unit: Depends on <expression> and the set unit.
Manual operation:
See
"Channel Calculation Function"
CALCulate<Measurement>:MATH[:EXPRession]:CATalog?
[<expressions>]
Queries all channel calculation functions supported by
.
Calculation functions