Remote control commands
R&S
®
NRX
215
User Manual 1178.5566.02 ─ 09
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
-1e18 to 1e18
*RST:
10.0
Default unit: -
Manual operation:
See
CALCulate<Measurement>:TRACe:Y[:SCALe]:TOP:WATT
<value>
Effective for trace, pulse analysis, time gate, timeslot measurements.
Sets the power reference value. The reference value is assigned to the top line of the
grid.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
-1e9 to 1e9
*RST:
1e-3
Default unit: W
Manual operation:
See
DISPlay[:WINDow<Window>]:ANALog:AUTO
<state>
DISPlay[:WINDow<Window>]:METer:AUTO
<state>
Automatically determines the scaling for the analog display. The upper and the lower
limit value are set depending on the current measurement data.
Suffix:
<Window>
.
1 to 4
Measurement pane
Parameters:
<state>
ONCE | OFF
*RST:
OFF
DISPlay[:WINDow<Window>]:ANALog:LOWer
<value>
DISPlay[:WINDow<Window>]:METer:LOWer
<value>
Sets the lower limit value of the analog scale.
Suffix:
<Window>
.
1 to 4
Measurement pane
Measurement settings and results