Remote control commands
R&S
®
NRX
397
User Manual 1178.5566.02 ─ 09
Example:
MEM:TABL:FREQ 50.00,60,70.3456
Sets 3 frequency values; 50.00
Hz, 60.00
Hz, 70.35
Hz.
Manual operation:
See
MEMory:TABLe:FREQuency:POINts?
Queries the number of frequency values in the selected correction table.
Use
Usage:
Query only
Manual operation:
See
MEMory:TABLe:GAIN:POINts?
Queries the number of offset values in the selected correction table.
Use
Usage:
Query only
Manual operation:
See
MEMory:TABLe:GAIN[:MAGNitude]
<value>...
Defines the offset values in the selected correction table. Existing data is overwritten.
Take care that the number of frequency values matches the number of offset values,
defined by
. If the numbers differ, excess values are
ignored.
Use
Parameters:
<value>
Numeric values with a maximum of 3
digits after the decimal
point, separated by commas. Values with more than 3
decimal
places are rounded.
Default unit: dB
Example:
MEM:TABL:GAIN 0,0.0033,0.04
Sets 3 offset values; 0.000
dB, 0.003
dB, 0.040
dB.
Manual operation:
See
MEMory:TABLe:MAP
<register_name>[, <register_number>]
Assigns a name to the correction table stored in the memory location.
Take care to use recognizable names. The R&S NRX does not check whether a name
is unique. If you want to reset the factory default, use
.
The query returns the memory location that is assigned to the setup name.
Managing setups and correction tables