![Rohde & Schwarz 1424.7805K02 User Manual Download Page 259](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791259.webp)
Remote control commands
R&S
®
NRX
259
User Manual 1178.5566.02 ─ 09
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Event
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?
Effective for continuous average measurements.
Queries the number of results that are currently stored in the result buffer.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:INFO?
[<ITEM>]
Effective for continuous average measurements.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Query parameters:
<ITEM>
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
<count>
Effective for continuous average measurements.
Sets the size of the result buffer.
You can enable the buffer using
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:
.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<count>
Range:
1 to 131072
*RST:
1