Measurement and display configuration
R&S
®
NRX
72
User Manual 1178.5566.02 ─ 09
Remote command:
Recall Parameter Set ← Quick Setup
Loads the parameters set selected under
Remote command:
6.2
Continuous average
The power sensor measures the signal average power asynchronously within a
defined time interval, the so-called aperture or sampling window. After a trigger event,
the power is integrated over the time interval.
The continuous average measurement type is the preferred measurement method if
the measurement is not to be, or cannot be, synchronized with a specific signal event.
It is the only available measurement type for thermal power sensors because they are
too slow for the other measurement types.
6.2.1
Continuous average result display
The measurement result is a single scalar value, either an absolute value or related to
a reference value.
Continuous average