Remote control commands
R&S
®
NRX
283
User Manual 1178.5566.02 ─ 09
CALCulate<Measurement>:TRACe:MEASurement:POWer:AVG?
Effective for pulse analysis measurements.
Queries the average power during the time the pulse is active.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation:
See
CALCulate<Measurement>:TRACe:MEASurement:POWer:HREFerence?
Effective for pulse analysis measurements.
Queries the power level at high reference level set by
TRACe:MEASurement:DEFine:TRANsition:HREFerence
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation:
See
CALCulate<Measurement>:TRACe:MEASurement:POWer:LREFerence?
Effective for pulse analysis measurements.
Queries the power level at low reference level set by
TRACe:MEASurement:DEFine:TRANsition:LREFerence
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation:
See
CALCulate<Measurement>:TRACe:MEASurement:POWer:MAX?
Effective for pulse analysis measurements.
Queries the maximum power measured within the analysis window.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Measurement settings and results