
Measurement and display configuration
R&S
®
NRX
88
User Manual 1178.5566.02 ─ 09
Pulse Base ← Power tab
Pulse base power level detected by the selected
. This value is used as a ref-
erence (0
%) to determine other parameter values such as the rising or falling thresh-
olds.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:POWer:PULSe:BASE[:
STATe]
CALCulate<Measurement>:TRACe:MEASurement:POWer:PULSe:BASE?
Trace Min ← Power tab
Minimum power measured within the analysis window.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:POWer:MINimum[:STATe]
CALCulate<Measurement>:TRACe:MEASurement:POWer:MIN?
High Ref. ← Power tab
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:POWer:PULSe:
HREFerence[:STATe]
CALCulate<Measurement>:TRACe:MEASurement:POWer:HREFerence?
Trace Avg ← Power tab
Average power during the time the pulse is active.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:POWer:AVG[:STATe]
CALCulate<Measurement>:TRACe:MEASurement:POWer:AVG?
Pos. Overshoot ← Power tab
Height of the local maximum before a falling edge, divided by the pulse amplitude:
Positive overshoot = 100 % x
Max. power - pulse top power
Pulse amplitude
.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:TRANsition:POSitive:
OVERshoot[:STATe]
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:POSitive:
OVERshoot?
Pulse analysis