Remote control commands
R&S
®
NRX
288
User Manual 1178.5566.02 ─ 09
Queries the height of the local maximum before a falling edge, divided by the pulse
amplitude:
Positive overshoot = 100 % x
Max. power - pulse top power
Pulse amplitude
DISPlay[:WINDow<Window>]:TRACe:
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation:
See
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:SPERiod?
Effective for pulse analysis measurements.
Queries the number of samples per second.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation:
See
DISPlay[:WINDow<Window>]:TRACe:MEASurement:POWer:AVG[:STATe]
<value>
Effective for pulse analysis measurements.
Enables or disables the display of the average signal power. See
CALCulate<Measurement>:TRACe:MEASurement:POWer:AVG?
Suffix:
<Window>
.
1 to 4
Measurement pane
Parameters:
<value>
*RST:
OFF
Manual operation:
See
DISPlay[:WINDow<Window>]:TRACe:MEASurement:POWer:MAXimum[:STATe]
<value>
Effective for pulse analysis measurements.
Measurement settings and results