Measurement and display configuration
R&S
®
NRX
59
User Manual 1178.5566.02 ─ 09
Off
Displays the absolute power or power ratio.
On
Displays the relative power or power ratio. As reference value, the
value specified under
Forward Reference Value, Reflection Refer-
Set
Assigns the current measurement result as reference value and dis-
plays the relative power.
Remote command:
CALCulate<Measurement>:RELative<DirectionalChannel>:STATe
Max Hold
If enabled, displays the highest value measured for each point (pixel) of the selected
display type.
Remote command:
CALCulate<Measurement>:HOLD[:STATe]
Max Hold Function
For all measurement functions, the R&S NRX stores the maximum and minimum val-
ues and the calculated differences between these values.
The selected setting applies to both power and reflection indication. You can change at
any time.
"Max"
Maximum value
"Min"
Minimum value
"Max
−
Min"
Difference between maximum and minimum value
Remote command:
CALCulate<Measurement>:HOLD:FUNCtion
Limit Monitor
Available for continuous average, burst average, time gate, timeslot, NRT measure-
ments.
For each window with digital or digital/analog result display, you can set an upper and
a lower limit.
Lower Limit State ← Limit Monitor
Available for continuous average, burst average, time gate, timeslot measurements.
Enables or disables the monitoring function for the lower limit.
Remote command:
CALCulate<Measurement>:LIMit<DirectionalChannel>:LOWer:STATe
Lower Limit ← Limit Monitor
Available for continuous average, burst average, time gate, timeslot measurements.
Available if
is set to "On".
Defines a lower limit.
Configuration for all measurement types