Queries the height of the local minimum before a rising edge, divided by the pulse
amplitude:
Negative overshoot = 100 % x
Pulse base power - minimum power
Pulse amplitude
DISPlay[:WINDow<Window>]:TRACe:
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation:
See
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:POSitive:
DURation?
Effective for pulse analysis measurements.
Queries the time that the pulse requires to transition from the pulse base level to the
pulse top level.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation:
See
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:POSitive:
OCCurrence?
Effective for pulse analysis measurements.
Queries the time when the signal passes through the medial reference power level with
rising edge, referenced to the delayed trigger event. Indicates the start point of the first
power pulse within the analysis window.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
Manual operation:
See
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:POSitive:
OVERshoot?
Effective for pulse analysis measurements.