Remote control commands
R&S
®
NRX
260
User Manual 1178.5566.02 ─ 09
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe
<state>
Enables or disables the buffered continuous average measurement. If enabled, all
results generated by trigger events are collected until the buffer is filled. Thus, a higher
data rate is achieved.
You can set the size of the buffer with
.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
ON | OFF
*RST:
OFF
13.6.7
Burst average measurements
There are several ways to programm the R&S NRX for measurements. See
ter 13.4, "Making measurements"
Further information:
●
13.6.7.1
Combining commands for burst average measurements
Parameter list for burst average measurements
Parameter
Description
<dtolerance>
Mandatory. Length of a time interval during that the power level can drop
below the trigger level without being interpreted as end of the power
pulse.
[SENSe<Sensor>:][POWer:]BURSt:DTOLerance
.
<start_exclude>
Mandatory. Amount of time at the beginning of a timeslot or integration
period that is not evaluated.
[SENSe<Sensor>:]TIMing:EXCLude:STARt
<end_exclude>
Mandatory. Amount of time at the end of a timeslot or integration period
that is not evaluated.
[SENSe<Sensor>:]TIMing:EXCLude:STOP
.
<expected_value>
Optional. Value that is expected for the measurement.
Measurement settings and results