![Rohde & Schwarz 1424.7805K02 User Manual Download Page 493](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791493.webp)
Remote control basics
R&S
®
NRX
493
User Manual 1178.5566.02 ─ 09
Bit no.
Short description
Bit is set if a measurement channel is waiting for trigger
event or for the extension summary.
7
Measurement m+7 wait for trigger
Channel
7
8
Measurement m+8 wait for trigger
Channel
8
14.4.8.6
Operation sense status register
The
CONDition
register contains information whether a power sensor is being initial-
ized. The
EVENt
register contains information whether an initialization was started or
completed since the last query.
A power sensor is initialized if:
●
Supply voltage is switched on (power-up).
●
Sensor was connected.
●
Reset was performed,
.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
Sensor A initializing
Sensor B initializing
Sensor C initializing
Sensor D initializing
NRT Sensor initializing
Other Sensor initializing
0
0
0
0
0
0
0
0
0
+
Figure 14-15: Operation sense status register
Querying the register:
●
STATus:OPERation:SENSe[:SUMMary]:CONDition?
●
STATus:OPERation:SENSe[:SUMMary][:EVENt]?
Table 14-17: Used operation sense status bits and their meaning
Bit no.
Short description
Bit is set if a sensor is initializing.
1
Sensor
A initializing
Sensor
A
2
Sensor
B initializing
Sensor
B
Status reporting system