Remote control commands
R&S
®
NRX
252
User Manual 1178.5566.02 ─ 09
Usage:
Query only
13.6.6.2
Combining continuous average commands with buffering
See also
Chapter 13.4, "Making measurements"
Parameter list for continuous average measurements with buffering
Parameter
Description
<buffered_size>
Mandatory. Number of requested measured values.
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
<expected_value>
Optional. Value that is expected for the measurement.
<resolution>
Optional. Limit up to which the measurement result probably is free of
noise.
[SENSe<Sensor>:]AVERage:COUNt:AUTO:
<source_list>
Mandatory. Primary and secondary sensor. The number for each sensor
is preceded by the character
@
. The entire expression is enclosed in
parentheses.
Example: (
@
3),(
@
2)
Sensor
C is the primary sensor, and sensor
B is the secondary sensor.
CONFigure<Measurement>:ARRay[:POWer][:AVG]?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
FETCh<Measurement>:ARRay[:POWer][:AVG]?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
READ<Measurement>:ARRay[:POWer][:AVG]?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
MEASure<Measurement>:ARRay[:POWer][:AVG]?
<buffered_size>[,
<expected_value_or_source_list>, <resolution_or_source_list>, <source_list>...]
Measured average power measured by one power sensor with buffering.
Suffix:
<Measurement>
.
1 to 8
Measurement
Query parameters:
<buffered_size>
<expr>
<expected_value_or_source_list>
<numeric> | <expr>
<resolution_or_source_list>
<numeric> | <expr>
<source_list>
<expr>
For the parameter descriptions, see
ous average measurements with buffering"
Usage:
Query only
Measurement settings and results