Remote control commands
R&S
®
NRX
377
User Manual 1178.5566.02 ─ 09
................................................................................................377
..................................................................................378
OUTPut:SOURce:STATe
<state>
SOURce:OUTPut:STATe
<state>
Requires the sensor check source (R&S NRX-B1)
Enables or disables the signal output.
Parameters:
<state>
*RST:
0
Manual operation:
See
SOURce:POWer[:VALue]
<value>
Requires the sensor check source (R&S NRX-B1)
Sets the power level for the output signal.
If you enter a value without unit, the unit is defined by
ther information, see
Parameters:
<value>
Range:
-40.0 to +20.0
*RST:
+0.0
Default unit: dBm
Manual operation:
See
SOURce:PULM:STATe
<state>
Requires the sensor check source (R&S NRX-B1)
Effective if the output signal is enabled, using
or
.
Sets the signal type of the output signal.
Parameters:
<state>
OFF | ON
OFF
Continuous wave
ON
Pulse modulation
*RST:
0
Manual operation:
See
Configuring the test generator