Remote control commands
R&S
®
NRX
447
User Manual 1178.5566.02 ─ 09
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:COUNt
<count>
Available for compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:POWer]:TSLot[:AVG]:COUNt
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<count>
Range:
1 to 128
*RST:
8
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:WIDTh
<width>
Available for compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:POWer]:TSLot[:AVG]:WIDTh
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<width>
Range:
50.0e-9 to 0.10
*RST:
1.000e-3
Default unit: s
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID:OFFSet[:TIME]
<time>
Available for compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:POWer]:TSLot[:AVG][:EXCLude]:MID:
OFFSet[:TIME]
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
0.00 to 0.10
*RST:
0.00
Default unit: s
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID:TIME
<time>
Available for compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:POWer]:TSLot[:AVG][:EXCLude]:MID:TIME
Remote emulation