
Remote control basics
R&S
®
NRX
494
User Manual 1178.5566.02 ─ 09
Bit no.
Short description
Bit is set if a sensor is initializing.
3
Sensor
C initializing
Sensor
C
4
Sensor
D initializing
Sensor
D
5
NRT sensor initializing
R&S
NRT
‑
Zxx power sensor
6
Other sensor initializing
USB or LAN power sensor
14.4.8.7
Operation lower limit fail status registers
The
CONDition
registers contain information whether a measured value is below a
configured lower limit. The
EVENt
registers contain information whether a measured
value dropped below a limit value since the last query.
For information on index
Chapter 14.4.1, "Hierarchy of the status registers"
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Extension Summary
Measurement m+1 Lower Limit Fail
Measurement m+2 Lower Limit Fail
Measurement m+3 Lower Limit Fail
Measurement m+4 Lower Limit Fail
Measurement m+5 Lower Limit Fail
Measurement m+6 Lower Limit Fail
Measurement m+7 Lower Limit Fail
Measurement m+8 Lower Limit Fail
0
0
0
0
0
0
0
+
Figure 14-16: Operation lower limit fail status registers
Querying the register:
●
STATus:OPERation:LLFail<RegisterIndex>[:SUMMary]:CONDition?
●
STATus:OPERation:LLFail<RegisterIndex>[:SUMMary][:EVENt]?
Status reporting system