Remote control commands
R&S
®
NRX
194
User Manual 1178.5566.02 ─ 09
INITiate<Measurement>:CONTinuous
<state>
Enables or disables the continuous measurement mode. In continuous measurement
mode, the power sensor does not go into the idle state after a measurement has been
completed, but immediately executes another measurement cycle.
The command is not fully R&S
NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S
NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<state>
instead of
ON
Measurements are performed continuously. If a measurement is
completed, the power sensor does not return to the idle state but
enters the waiting for trigger state again.
OFF
Ends the continuous measurement mode, and sets the power
sensor to the idle state.
*RST:
0
INITiate<Undef>:ALL:CONTinuous
<state>
Applies to all connected power sensors. See
Suffix:
<Undef>
.
1 to n
No suffix required.
Setting parameters:
<state>
*RST:
0
Usage:
Setting only
INITiate<Measurement>:DISable
<state>
Prevents the execution of
INITiate<Measurement>[:IMMediate]
prevent that the specified power sensor starts a measurement if
is used.
The command is not fully R&S
NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S
NRP2 emulation using
"NRP2"
.
Starting and ending a measurement