![Rohde & Schwarz 1424.7805K02 User Manual Download Page 295](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791295.webp)
Remote control commands
R&S
®
NRX
295
User Manual 1178.5566.02 ─ 09
13.6.10
Time gate measurements
Further information:
●
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG]:OFFSet[:TIME]
.......................... 295
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG]:TIME
........................................ 295
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID:OFFSet[:TIME]
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID:TIME
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID[:STATe]
CALCulate<Measurement>[:POWer]:TGATe<Undef>[:AVG]:SELection
................................297
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG]:OFFSet[:TIME]
<value>
Sets the length of the gate.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Gate>
1 to 4
Time gate
Parameters:
<value>
Range:
0.0 to 15.0
*RST:
0.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG]:TIME
<value>
Sets the start time of the gate.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Gate>
1 to 4
Time gate
Parameters:
<value>
Range:
50.0e-9 to 0.1
*RST:
1.0e-3
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID:
OFFSet[:TIME]
<value>
Sets length of the fence.
Measurement settings and results