Remote control commands
R&S
®
NRX
195
User Manual 1178.5566.02 ─ 09
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<state>
*RST:
0
INITiate<Undef>:ALL:DISable
<state>
Applies to all connected power sensors. See
Suffix:
<Undef>
.
1 to n
No suffix required.
Setting parameters:
<state>
Usage:
Setting only
INITiate<Measurement>[:IMMediate]
Starts a single measurement cycle. The power sensor changes from the idle state to
the waiting for trigger state. As soon as the trigger condition is fulfilled, the sensor
begins the measurement. Depending on the number of trigger events that are required,
e.g. for averaging, the power sensor enters the waiting for trigger state several times.
Once the entire measurement is completed, a measurement result is available, and the
power sensor enters the idle state again.
Use the command only after the continuous measurement mode has been disabled
using
INITiate<Measurement>:CONTinuous
OFF
.
The command is not fully R&S
NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S
NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Example:
If you work in a sender/receiver setup, you need to trigger the
receiver before triggering the sender. To prevent overlapping
execution, use
*WAI
sensor
2 is the receiver, sensor
1 is the sender:
INIT2
*WAI
INIT1
Usage:
Event
Starting and ending a measurement